Cell culture confirmation of respiratory specimens from frozen repositories and fresh isolates were collected and tested at three separate geographical locations in the United States.
ViraSTAT® and another in-vitro diagnostic kit were compared. A total of 278 specimens were examined with the FITC-labeled Anti-Parainfluenza Pool and a total of 434 specimens were examined with the FITC-labeled monoclonal antibody identification reagents. |
 |
 |
 |
 |
 |
 |
 |
 |
| Para Pool |
Para 1 |
Para 2 |
Para 3 |
|
 |
| 80 specimens |
52 specimens |
18 specimens |
101 specimens |
Total Positive |
 |
| 100% |
98.1%1 |
100% |
100% |
Sensitivity |
 |
| 100% |
100% |
100% |
99.7%2 |
Specificity |
 |
1 One specimen was weakly positive with the reference comparison antibody and negative with ViraSTAT and negative with a confirming independent antibody. However, the reference antibody was considered the comparison standard.
2 One specimen was weakly negative with the reference comparison antibody but strongly positive with ViraSTAT and positive with a confirming independent antibody. However, the reference antibody was considered the comparison standard. |
|
 |
|